Geophysical Survey Systems, Inc. HandySan
User’s Manual
MN70-380 Rev B 2
HandyScan Operation
1. Install a charged battery.
2. Turn power On (switch is located on the back panel).
3. Wait until data screen appears (about 15 sec).
4. Check battery indicator (top right corner of the screen).
5. If a low battery is indicated, turn system off and replace.
Using Setup Table:
1.
Press Set on the keypad to display setup table, then:
Up/down arrows to select a parameter,
Left/right arrows to change it.
Disp. Direction: normal / invert
Disp. Mode : B / AB
Amplitude : abs / offset
X-axis : distance / time
Depth : ns / -3 to 3
Contrast : (number)
Date/time : 05/10/2001
Data_No. : 1
Dist adj. : +0 [0.00m]
Output : printer / computer
Baud rate : 57600
Default : NO
Data Collection
1. Press and hold down Start button, you will hear a single beep that denotes that the system is
ready for data collection.
2. Move unit forward.
3. Slow down if signal sounds when moving.
4. Release Start button and press Mark to insert a mark into data, press and hold Start button
to continue collecting data. If the Start button is not depressed within ten seconds, the unit
will close the data file.
5. Release the Start button and press the Down arrow to close file (double beep).